Universal Indicator with Retransmission Option

Product Code: PR40

The PR40 universal indicator and alarm controller offers a one-product solution for hundreds of applications!

Can be supplied pre-configured to suit your application requirements or easily configured with the TX-USB configuration kit.

Available in two versions: PR40U (indicator only) or PR40UR (indicator with 2 relay outputs and retransmission). The retransmission output option enables the indicator to also be used as a transmitter.

  • Features
  • Technical
  • Documents
  • Software
    • Universal input
    • Easy USB setup in minutes
    • USB powered setup
    • Pre-calibrated ranges
    • Designed for harsh environments
    • Large 20mm high super-bright LED display
    • Relay/Retransmission options
    • 20V excitation (to power transmitters)
    • Shallow depth unit
    • Custom linearisation
    • 0.03% basic accuracy
  • Technical Details (Common)
    Power Supply 100-300V DC / 85-265V AC
    Input Universal (see below for input specifications)
    EMC Compliance Emissions (EN 61326), Immunity (EN 61326), Safety (EN 61010-1)
    Display Super bright 20mm (0.8 inch) display 4 digits (7 segment)
    Accuracy  < ±0.03% FSO typical
    Ambient Drift < ±0.003%/°C FSO typical
    Noise Immunity 125dB CMRR average (2.0kV DC limit)
    R.F. Immunity <1% effect FSO typical
    Response Time 400ms (RTD / Thermocouple), 100ms other ranges
    Isolation Test Voltages Between input/output 3750V AC for 1 minute
    Operating Temperature -10°C to +50°C
    Storage Temperature -20°C to +60°C
    Operating Humidity 5-85% RH max (non-condensing)
    Dimensions 96mm (W) x 48mm (H) x 53mm (D) (panel mounting)
    Panel Cutout Dimensions 92mm (W) x 45mm (H)

     

    RTD Input Specifications
    Input Pt100 or Pt1000 DIN 3-wire type (2-wire can be used with offset calibration)
    Sensor Current 0.3mA nominal
    Sensor Break Output Drive Function high upscale/low downscale
    Lead Wire Resistance Pt100 (10Ω/wire max), Pt1000 (5Ω/wire max) 0.02% FSO offset error per Ω of lead resistance mismatch
    Linearity (Pt100) 0.02% FSO for span inputs ≤ +200°C
    0.1% FSO for span inputs ≤+300°C
    Linearity (Pt1000) 0.02% FSO for span inputs ≤ +200°C
    0.2% FSO for span inputs ≤+300°C

     

    Thermocouple Input Specifications
    Thermocouple Types K, J, T, N, E, R, S, B
    Input Impedance 1MΩ min
    Thermocouple Lead Resistance 100Ω max
    Cold Junction Compensation -10°C to +70°C
    Accuracy Types K, J, T, N, E: < ±1°C
    Types R, S, B: < ±2°C
    Temperature Drift Types K, J, T, N, E: < ±0.05°C/C
    Types R, S, B: < ±0.2°C/C
    Sensor Break Output Drive Function high upscale/low downscale
    CJC Error < ±1°C

     

    N.T.C. Input Specifications
    OEM Customers - contact sales for further information

     

    Current Input Specifications
    USB Programmable Zero 0 to ±99% of span
    Field Programmable Span 1μA-24mA DC
    Input Resistance 10Ω
    Maximum Over-Range 50mA DC continuous
    Linearity and Repeatability < ±0.02% FSO typical

     

    Voltage Input Specifications
    Ranges 0-1V, 0-10V, +10V to -10V, 0-30V, 0-300V
    USB Programmable Zero 0 to ±99% of span
    USB Programmable Span 95% of FSO
    Input Resistance 1MΩ min
    Linearity and Repeatability < ±0.02% FSO typical

     

    Potentiometer Input Specifications
    Potentiometer Input 3-wire
    Excitation Voltage Variable
    Potentiometer Resistance 0-2kΩ low pot, 0-5kΩ med pot, 0-20kΩ high pot
    Field Programmable Zero 0-90% of span
    Field Programmable Span 0.1-100%
    Linearity and Repeatability < ±0.05% FSO typical

     

    mV Input Specifications
    Ranges 0-200mV, ±200mV
    USB Programmable Zero 0 to ±99% of span
    USB Programmable Span 95% of FSO
    Input Resistance 1MΩ min
    Linearity and Repeatability < ±0.02% FSO typical

     

    Digital Input Specifications
    Flow Meter (pulse) 0-2000.0Hz open collector
    RPM Meter (pulse) 0-2000.0Hz open collector

     

    Relay Output Specifications
    2 x Form A Relays 250V AC, 3A (display includes 2 x setpoint indicator LEDs)

     

    Retransmission Specifications
    Current Output 4~20mA or 20~4mA
    20mA max output drive (600Ω max load at 12V DC)
    Resolution 1μA
    Accuracy/Repeatability 0.05% of FSO
    Temperature Drift 30ppm/°C typical

     